Instrument dimensions

 

Recommended Installation Space: W 76.78" × D 40"
Note: Leave at least 13.78" of space behind the CT system table and 6" of space left side in order to permit access for maintenance work. 

XSeeker and the Analytical Intelligence logo are trademarks of Shimadzu Corporation or its affiliated companies in Japan and/or other countries.
VGSTUDIO MAX is a trademark of Volume Graphics GmbH.

 

X-Ray Generator Output

160 kV 1.2 mA

X-Ray Detector

Flat panel detector

Max. Detector Input Resolution

Approx. 5.6 million pixels

Max. Sample Size/Weight

φ11.8"×H 12.6",22 lbs (Including sample setting jig)

Max. CT-Scan Region

SRD = 18.7" : φ3.34" × H 2.8"
SRD = 21.7" : φ3.9"×H 3.2"

Max. System Resolution Capability

100 µm

CT Data Acquisition Time

12 sec or 5 minutes

System Rated Output

Single phase 100 to 240 V AC ± 10 %, 50/60 Hz, 1.0 kVA

Rated Output : Control Computer

Single phase 100 to 240 V AC ± 10 %, 50/60 Hz, 1.0 kVA

Rated Output : Grounding

D type grounding (Grounding resistance of 100 Ω max.)

Weight

639 lbs

Operating Environment Conditions

Ambient temperature: 10 to 30 °C; Ambient humidity: RH of 45 to 80 % max. (No condensation)

External Leakage Dose

1 μSv/h or less.

Minimum Focal Spot Size

0.003"

Read more
Show Less
{"title":"Specification Sheet Downloads","description":"Download the latest specification sheets.","source":"product","key":4662,"max":30,"filter_types":["spec_sheet"],"link_title":"View other Downloads","link_url":""} {"title":"Downloads","description":"Download the latest brochure.","source":"product","key":4662,"max":30,"filter_types":["brochures"],"link_title":"View other Downloads","link_url":"","pdf_links":[]}