Nexera Preparative LC System - Options
Preparative Purification Liquid Chromatograph
The LH-40 Liquid Handler, Combination of Autosampler and Fraction Collector
Options
Syringe Kit 20 mL
This kit enables large-capacity injections of 2 mL or more at one time. The maximum injection volume is 20 mL.
Washing Pump
This reduces the washing time for the injection needle, increasing throughput while reducing carryover.
Analysis Kit
The recovered fraction can be reanalyzed to check the purity.
Liquid Surface Detection Needle
This detects the liquid surface level, and automatically determines whether there is any sample present. As a result, only the remaining volume is injected, which prevents the injection of air into columns. Additionally, if no sample is present, the system can proceed to the next sample, reducing needless errors and lost labor.
Multi-Liquid Handler Kit*1
Up to six LH-40 liquid handlers can be connected, making it easy to inject the sample from all LH-40. *1 Up to one LH-40 when FRC-40 fraction collectors are connected.
Autosampler
SIL-10AP
Sample Racks
Sample Coolers
Block Cooling/Heating: 4 to 70 °C)
Manual Injecto
Rheodyne®7725
Optional Sample Loops (Material: SUS)
Volume | Part Number(P/N) |
---|---|
100 μL | 228-32211-16 |
200 μL | 228-32211-17 |
500 μL | 228-32211-18 |
1 mL | 228-32211-19 |
5 mL | 228-32211-21 |
Specifications
LH-40 | SIL-10AP | |
---|---|---|
Maximum Injection Volume | Up to 2 mL, up to 20 mL (Optional available) | Up to 5 mL |
Cooling Function | None | Yes (Optionally available) |
Compatible Containers and Quantities |
10 mmO.D. test tube 540 pcs 1.5 mL vial 486 pcs*2
12 mmO.D. test tube 486 pcs 4.0 mL vial 252 pcs
13 mmO.D. test tube 360 pcs 13 mL vial 108 pcs
15 mmO.D. test tube 252 pcs 50 mL vial 54 pcs
17, 18 mmO.D. test tube 216 pcs 250 mL bottle 20 pcs
25 mmO.D. test tube 108 pcs 500, 1000 mL bottle 12 pcs
30, 35 mmO.D. test tube 54 pcs 96 well MTP/DWP 9 pcs
|
1.5 mL vials 100 pcs
4.0 mL vials 80 pcs
13 mL vials 25 pcs
|
Compatible Containers and Quantities | W 390 x H 560 x D 730 mm, 40 kg | W 260 x H 280 x D 320 mm, 18.5 kg |
*2 Available for injection sample container. Not available for fraction container.
FRC-40, Highly Flexible Fraction Collector
Sample Racks
A variety of containers can be placed including MTPs, vials, and various types of test tubes. Six colors are available, so a separate color can be apportioned to each user in order to avoid confusing samples.
Multi Fraction Collector Kit
Up to six FRC-40 fraction collectors (up to five LH-40 liquid handlers) can be connected, making it easy to increase the number of fractions.
Compact Design for Small-Volume Samples
FRC-10A
For smaller scale collection, or specialized applications that require enclosure and cooling, the FRC-10A is a compact fraction collector that provides time and signal based triggering. A variety of programmable fractionation functions enable target components to be collected with high purity and high recovery.。
Specifications
FRC-40 | FRC-10A | |
---|---|---|
Maximum Number of Fractions | Up to 540 pcs* | Up to 144 pcs* |
Maximum Flow Rate | 150 mL/min | |
Fractionation Mode | Configured through a combination of basic mode (initial parameter mode) and time program mode (14 parameters) | |
Cooling Function | None | Optionally available |
Compatible Containers and Quantities |
10 mmO.D. test tube 540 pcs 4 mL vial 252 pcs
12 mmO.D. test tube 486 pcs 15 mL vial 108 pcs 16 mmO.D. test tube 216 pcs 50 mL vial 54 pcs 18 mmO.D. test tube 216 pcs 250 mL vial 20 pcs 25 mmO.D. test tube 108 pcs 500/1000 mL bottle 12 pcs 35 mmO.D. test tube 54 pcs MTP/DWP 9 pcs |
10 mmO.D. test tubes 144 pcs
18 mmO.D. test tubes 164 pcs 35 mmO.D. test tubes 16 pcs |
Size and weight | W390 × H560 × D730 mm, 30 kg | W 260 x H 280 x D 420 mm,15 kg |
Example of Scaling Up from Analytical to Semi Preparative Work
This is an example of scaling up in which seven types of parabens are targeted using a 150 mm long column with a particle size of 5 μm.
The gradient elution conditions investigated at the analytical scale are transitioned to the semi-preparative scale. A comparable chromatogram is obtained at both scales.
Column: Shim-pack Scepter C18-120 (4.6 mm x 150 mm,5 μm)
Column: Shim-pack Scepter C18-120 (20 mm x 150 mm,5 μm)
Suited to the Target Preparative Method
Column Hub
Column Holder
Column Holder SLIM
Preparative columns with an I.D. of 20 mm to 50 mm as well as manual switching valves can be attached. The valves can be used for column switching.
Specifications
Installable Valves | Installable Columns | Size | |
---|---|---|---|
Column Hub | Automatic Switching Valves Up to 4 pcs | Preparative Columns 2 pcs Analytical Columns 6 pcs* | W 260 x H 560 x D 500 mm |
Column Holder | Manual Switching Valves Up to 4 pcs | Preparative Columns 2 pcs Analytical Columns 1 pc | W 250 x H 465 x D 400 mm |
Column Holder SLIM | Manual Switching Valves Up to 5 pcs | Preparative Columns 2 pcs Analytical Columns 1 pc** | W 110 x H 625 x D 500 mm |
* When attaching three or more analytical columns or two preparative columns to the column hub, the optional "Column Bracket" is needed.
** If two preparative columns are attached using the column holder SLIM, the optional column clamp ASSY (P/N: 228-17701-94) is required.
For Multiple Detection Triggers
A/D Conversion Board Kit
This is required for preparative work using multiple detector triggers. Expand the hardware to suit the number of detection trigger channels required.
Degassing Units
DGU-403 / DGU-405
- A low-capacity degassing unit that uses a special fluororesin membrane.
DGU-403 : 3 flow lines, DGU-405 : 5 flow lines - The maximum operating flow rate per flow line is 10 mL/min.
- Designed for use in analytical and preparative fractionation, this unit is used only when retention time reproducibility needs to be improved during analysis.
* When connecting to an LC-20AP, a connection kit must be obtained separately.
* LC-20AR connection kit is required when the operating flow rate is more than 10 mL/min.
Helium Degassing Unit
DGU-10B
- Eliminates air bubbles, baseline undulation, drifting, etc. by purging dissolved air from mobile phases.
- The DGU-10B can be used to degas up to four mobile phase solutions with helium gas.
- This unit is switched ON/OFF from the solvent delivery unit or system controller.
High-Pressure Flow-Line Selection Valves
FCV-20AH2 / FCV-12AH
- The valve position is controlled by event signal input.
- Valve type: 2-position/6-port rotary valve (recycle valve : 2-position/3-port valve)
- Maximum operating pressure: 34.3 MPa
- Operating pH range: pH 1 to 10
- Operating temperature range: 4 to 35°C
- Storing the FCV-12AH in the Option Box helps reduce the volume of preparative piping, including the recycling flow lines.
Low-Pressure Gradient Unit
FCV-200AL
- This low-pressure gradient unit is for the LC-20AP large-volume solvent delivery pump.
- A gradient can be produced with a single pump, enabling gradient preparative work at low cost.
- A single unit is capable of providing up to a four-liquid gradient as well as solvent switching, reducing the work involved in mobile phase investigations during method development.
Reservoir Selection Valves
FCV-11AL / FCV-11ALS / FCV-230AL
- Capable of switching solvents using solenoid valves.
- The FCV-11AL/FCV-11ALS provide switching between two solvents. The FCV-11AL can supply up to three solvent delivery units, whereas the FCV-11ALS is used for one unit. It can be controlled from the LC-20AP/20AR front panel directly or through a system controller CBM-20A/20Alite and workstation software.
- The FCV-230AL provides switching between two solvents (optionally four solvents). It can be controlled from the LC-20AP/20AR front panel directly or through a system controller CBM-20A/20Alite and workstation software.
Related information
-
The ultra high-speed LC/MS Prominence UFLC+LCMS-2020 improves the speed and reliability of HPLC analysis, offering dramatic improvements in lab productivity.
-
Efficient, Easy to Use, Space Saving
Up to six fraction collectors can be connected simultaneously.
Compatible with many racks corresponding to the intended fraction volume.
Can accommodate up to 3,240 fractions (6 FRC-40s, 10 mm tubes).
-
The Nexera series is a family of HPLC systems that marries AI and IoT enhancements to set new industry standards.