Analysis of lipids in full rat brain with 30 μm and 50 μm spacing
MALDI images of full rat brain using FlexiVision-mini ITO glass slide. A, optical scan; B, positive ion mode image from MALDI-8020 using 50 μm spacing; C, negative ion mode image from MALDI-8030 using 30 μm spacing.
Sample B
Matrix | DHB, sublimated with Shimadzu iMLayer device |
Measurement region | 67,024 pixels |
Measurement time | around 1.9 hours |
Experiment details | Laser repetition rate of 200 Hz and 20 shots per pixel |
Sample C
Matrix | 9-AA, sublimated with Shimadzu iMLayer device |
Measurement region | 275,210 pixels |
Measurement time | around 3.8 hours |
Experiment details | Laser repetition rate of 200 Hz and 10 shots per pixel |
Large molecule imaging (protein/ on-tissue digestion) with 50 μm spacing
MALDI images of protein and peptides in full rat brain using FlexiVision-mini ITO glass slide and MALDI-8020. A, MBP protein, (m/z 14124); B, MBP peptide (HGFLPR, m/z 726.39)
Sample A
Matrix | Sinapinic acid, sprayed with Shimadzu iMLayer AEROdevice |
Measurement region | 68,836 pixels |
Measurement time | around 3.8 hours |
Experiment details | Laser repetition rate of 100 Hz and 20 shots per pixel |
Sample B
Matrix | CHCA, sprayed with Shimadzu iMLayer AERO device |
Measurement region | 67,482 pixels |
Measurement time | around 2.8 hours |
Experiment details | Laser repetition rate of 200 Hz and 30 shots per pixel |