Nexera Analytical SFC - Specs
Supercritical Fluid Chromatography System
Special Units
Supercritical Fluid Extraction Unit SFE-30A
This unit operates at a maximum temperature of 80°C to allow faster and more complete extractions. Two sizes of extraction vessels (5 mL and 0.2 mL) can be chosen based on the sample amount. This unit has the internal capacity to run up to four samples automatically; with the addition of an optional rack changer, it can accommodate unattended operation for up to 48 samples.
CO2 Delivery Unit LC-30ADSF/Backpressure Control Unit SFC-30A
Stable baseline with low pulsation is realized by LC-30 ADSF with a built-in cooler for pump heads, which delivers mobile phases up to 5 mL/min at pressures up to 66 MPa. The low dead volume of SFC-30A (0.7 uL) allows a mass spectrometer to be directly connected to the SFC system with out splitting so that higher sensitivity can be achieved.
Nexera UC reduces environmental impact
Comparison of costs and consumption of organic solvent for a single analysis by conventional normal-phase LC vs. SFC is shown below. By using SFC, the total cost of analysis is reduced by 87.6% and the consumption of organic solvent is reduced by 94.2%.