Nexera XS inert UHPLC System - Specs
Ultra High Performance Liquid Chromatograph
Allowable maximum pressure | 105 MPa |
---|---|
Internal rinsing | Standard equipment (3 dil) |
Features | Nexera XS inert combines excellent resistance to corrosion with the inertness of surfaces in contact with the sample and mobile phases. The system provides excellent chromatographic separation performances for a wide range of applications. |
Model Name | Features | |
---|---|---|
System controller | SCL-40* CBM-40* CBM-40lite* |
Color screen with touch panel and system controller with reservoir tray Number of connectable units : 8 (Using option: 12) , 4(CBM-lite, Excluding built-in solvent delivery unit) Connectable unit : Solvent delivery unit: max. 4, Autosampler: 1, Column oven: max. 4, Detector: max 2, etc. |
Solvent Delivery Pump | LC-40D XSi | Pumping method : Parallel-type double plunger (approx. 10 μL/1 stroke) Allowable maximum pressure : 105 MPa Flow rate settings range : 0.0001 to 3.0000 mL/min (1.0 to 105 MPa), 3.0001 to 5.0000 mL/min (1.0 to 80 MPa), 5.0001 to 10.0000 mL/min (1.0 to 22 MPa) Flow rate accuracy :≦± 1% (under specified conditions) Gradient mode : High-pressure gradient (2 or 3 solvents) Quaternary low-pressure gradient |
Gradient Mixer | MR Mixer(Inert) | MR 20 µL MR 180 µL MR 40 µL LPGE MR 300 µL LPGE |
Degassing Unit | DGU-403* DGU-405* |
Number of degassed solvents : 3(DGU-403), 5(DGU-405) Degassed low line capacity : 400 μL/1 line Operating temperature range : 4 to 35°C |
Autosampler | SIL-40C XSi | Injection method : Total-volume Injection (standard), loop injection (optional) Allowable maximum pressure : 105 MPa Injection volume : 0.1 to 50 μL, 0.1 to 2000 μL (optional) Injection volume accuracy : ≦± 1% (5 μL injection, n = 20) Injection cycle time : ≦6.7 seconds (under specified conditions) |
Column Oven | CTO-40S | Temperature control type : Forced air circulation Cooling Method : Electronic cooling Temperature control range : Room temperature -10 to 85 ℃(CTO-40S) Room temperature -10 to 100 ℃(CTO-40C) Temperature accuracy :±0.8℃(CTO-40S), ±0.5℃(CTO-40C) Temperature precision : 0.2 ℃(±0.1 ℃、CTO-40S), 0.1 ℃(±0.05 ℃、CTO-40C) |
UV-VIS Detector | SPD-M40* Please contact your local Shimadzu representative. |
Light source : Deuterium (D2) lamp Number of diode elements : 1024 Wavelength range : 190 to 700 nm Wavelength accuracy : ≦± 1 nm Wavelength reproducibility : ≦± 0.1 nm |
Flow Cell | Flow Cell | Cell volume: 8 μL, Cell path length: 10 mm |
Low-diffusion inert cell | Cell volume: 2.5 μL, Cell path length: 5 mm | |
Inert cell | Cell volume: 12 μL, Cell path length: 10 mm | |
pH monitor | pHM-40 | Measurement range:pH 1 to 14 Measurement error:pH ± 0.1 Ambient operating temperature:4 to 35 ℃ Usable humidity range:20 to 85 % |
Software | Labsolutions LC/GC Multi LC-PDA, LabSolutions LCMS Please contact your local Shimadzu representative. |