AP225W

Capacity

220g

mol Conversion Function

Included

Display

OEL display (dot matrix)

Weight

Approx. 7.9 kg

Body Dimensions

Approx. 212(W) x 411(D) x 345(H)mm
(power supply unit included)

Pan Size

ø91 mm

Sensitivity Stability Against Temperature Range

±2 ppm/°C (10 to 30°C)

Operating Temperature/Humidity Range

5 to 40°C 20 to 85% *5

Ionizer

Optional

Clock-CAL

Included

Inspection Support Function

Included

Sample (Concentration) Preparation

Included

HPLC Buffer Solution Preparation

Included

Minimum display

0.01 mg

Recipe Compounding

Included

USB Device (Type B)

Included

USB Host (Type A)

Included

Response Time *4 

8 sec.

Response Time for race Measurements *3

2 sec.

Linearity *2

±0.1 mg

Minimum Weight *2

20 mg

Repeatability (Standard deviation)

0.015mg
(to 20g)
0.03mg
(to 100g)
0.05mg
(to Minimum Weight)

External Calibration Weight Range for Span Calibration

95 to 220.00090 g
(200g)

Calibration Weight

Built-in *1

Input/Output Terminal

RS-232C (D-sub9P plug) USB host (Type A) USB device (Type B) Ionizer

*1 Minimum display 0.01mg models provide two internal weights as standard.
*2 Be compliant with USP Chapter 41. This is the tested value by the weight of the balance’s capacity of 5% (or 5 grams’ weight). In the case of the AP225W, the results are for tests carried out with the optional internal windbreak plate applied. The minimum weight value is affected by the installation environment, so it is necessary to measure it in the actual environment of use.
*3 The response time for displaying 90% of the added sample amount value in trace measurements (from 1mg)
*4 The response time value is typical.
*5 Non-condensing.

Read more
Show Less

AP125WD

Capacity

120 g / 52g

mol Conversion Function

Included

Display

OEL display (dot matrix)

Weight

Approx. 7.9 kg

Body Dimensions

Approx. 212(W) x 411(D) x 345(H)mm
(power supply unit included)

Pan Size

ø91 mm

Sensitivity Stability Against Temperature Range

±2 ppm/°C (10 to 30°C)

Operating Temperature/Humidity Range

5 to 40°C 20 to 85% *5

Ionizer

Optional

Clock-CAL

Included

Inspection Support Function

Included

Sample (Concentration) Preparation

Included

HPLC Buffer Solution Preparation

Included

Minimum display

0.01 mg / 0.1 mg

Recipe Compounding

Included

USB Device (Type B)

Included

USB Host (Type A)

Included

Response Time *4 

2 sec. / 8 sec

Response Time for race Measurements *3

2 sec.

Linearity *2

±0.2 mg /±0.05 mg

Minimum Weight *2

30 mg

Repeatability (Standard deviation)

0.1 mg / 0.02 mg

External Calibration Weight Range for Span Calibration

95 to 120.0090 g
(100 g)

Calibration Weight

Built-in *1

Input/Output Terminal

RS-232C (D-sub9P plug) USB host (Type A) USB device (Type B) Ionizer

Read more
Show Less

AP225WD

Capacity

220 g / 102g

mol Conversion Function

Included

Display

OEL display (dot matrix)

Weight

Approx. 7.9 kg

Body Dimensions

Approx. 212(W) x 411(D) x 345(H)mm
(power supply unit included)

Pan Size

ø91 mm

Sensitivity Stability Against Temperature Range

±2 ppm/°C (10 to 30°C)

Operating Temperature/Humidity Range

5 to 40°C 20 to 85% *5

Ionizer

Optional

Clock-CAL

Included

Inspection Support Function

Included

Sample (Concentration) Preparation

Included

HPLC Buffer Solution Preparation

Included

Minimum display

0.01 mg / 0.1 mg

Recipe Compounding

Included

USB Device (Type B)

Included

USB Host (Type A)

Included

Response Time *4 

2 sec. / 8 sec

Response Time for race Measurements *3

2 sec.

Linearity *2

±0.2 mg /±0.1 mg

Minimum Weight *2

30 mg

Repeatability (Standard deviation)

0.1 mg / 0.05 mg

External Calibration Weight Range for Span Calibration

95 to 220.009 g
(200 g)

Calibration Weight

Built-in *1

Input/Output Terminal

RS-232C (D-sub9P plug) USB host (Type A) USB device (Type B) Ionizer

Read more
Show Less

AP124W

Capacity

120 g

mol Conversion Function

Included

Display

OEL display (dot matrix)

Weight

Approx. 7.0 kg

Body Dimensions

Approx. 212(W) x 367(D) x 345(H)mm

Pan Size

ø91 mm

Sensitivity Stability Against Temperature Range

±2 ppm/°C (10 to 30°C)

Operating Temperature/Humidity Range

5 to 40°C 20 to 85% *5

Ionizer

Optional

Clock-CAL

Included

Inspection Support Function

Included

Sample (Concentration) Preparation

Included

HPLC Buffer Solution Preparation

Included

Minimum display

0.1 mg

Recipe Compounding

Included

USB Device (Type B)

Included

USB Host (Type A)

Included

Response Time *4 

2 sec.

Response Time for race Measurements *3

2 sec.

Linearity *2

±0.2 mg

Minimum Weight *2

200 mg

Repeatability (Standard deviation)

0.1 mg

External Calibration Weight Range for Span Calibration

45 to 120.009 g
(100 g)

Calibration Weight

Built-in *1

Input/Output Terminal

RS-232C (D-sub9P plug) USB host (Type A) USB device (Type B) Ionizer

Read more
Show Less

AP224W

Capacity

220 g

mol Conversion Function

Included

Display

OEL display (dot matrix)

Weight

Approx. 7.0 kg

Body Dimensions

Approx. 212(W) x 367(D) x 345(H)mm

Pan Size

ø91 mm

Sensitivity Stability Against Temperature Range

±2 ppm/°C (10 to 30°C)

Operating Temperature/Humidity Range

5 to 40°C 20 to 85% *5

Ionizer

Optional

Clock-CAL

Included

Inspection Support Function

Included

Sample (Concentration) Preparation

Included

HPLC Buffer Solution Preparation

Included

Minimum display

0.1 mg

Recipe Compounding

Included

USB Device (Type B)

Included

USB Host (Type A)

Included

Response Time *4 

2 sec.

Response Time for race Measurements *3

2 sec.

Linearity *2

±0.2 mg

Minimum Weight *2

200 mg

Repeatability (Standard deviation)

0.1 mg

External Calibration Weight Range for Span Calibration

95 to 220.009 g
(200 g)

Calibration Weight

Built-in *1

Input/Output Terminal

RS-232C (D-sub9P plug) USB host (Type A) USB device (Type B) Ionizer

Read more
Show Less

AP324W

Capacity

320 g

mol Conversion Function

Included

Display

OEL display (dot matrix)

Weight

Approx. 7.0 kg

Body Dimensions

Approx. 212(W) x 367(D) x 345(H)mm

Pan Size

ø91 mm

Sensitivity Stability Against Temperature Range

±2 ppm/°C (10 to 30°C)

Operating Temperature/Humidity Range

5 to 40°C 20 to 85% *5

Ionizer

Optional

Clock-CAL

Included

Inspection Support Function

Included

Sample (Concentration) Preparation

Included

HPLC Buffer Solution Preparation

Included

Minimum display

0.1 mg

Recipe Compounding

Included

USB Device (Type B)

Included

USB Host (Type A)

Included

Response Time *4 

2 sec.

Response Time for race Measurements *3

2 sec.

Linearity *2

±0.3 mg

Minimum Weight *2

200 mg

Repeatability (Standard deviation)

0.15 mg

External Calibration Weight Range for Span Calibration

95 to 320.009 g
(300 g)

Calibration Weight

Built-in *1

Input/Output Terminal

RS-232C (D-sub9P plug) USB host (Type A) USB device (Type B) Ionizer

Read more
Show Less

AP124X

Capacity

120 g

mol Conversion Function

Included

Display

OEL display (dot matrix)

Weight

Approx. 7.0 kg

Body Dimensions

Approx. 212(W) x 367(D) x 345(H)mm

Pan Size

ø91 mm

Sensitivity Stability Against Temperature Range

±2 ppm/°C (10 to 30°C)

Operating Temperature/Humidity Range

5 to 40°C 20 to 85% *5

Ionizer

Optional

Clock-CAL

Included

Inspection Support Function

Included

Sample (Concentration) Preparation

Not Included

HPLC Buffer Solution Preparation

Not Included

Minimum display

0.1 mg

Recipe Compounding

Not Included

USB Device (Type B)

Included

USB Host (Type A)

Not Included

Response Time *4 

2 sec.

Response Time for race Measurements *3

2 sec.

Linearity *2

±0.2 mg

Minimum Weight *2

200 mg

Repeatability (Standard deviation)

0.1 mg

External Calibration Weight Range for Span Calibration

45 to 120.009 g
(100 g)

Calibration Weight

Built-in

Input/Output Terminal

RS-232C (D-sub9P plug)
USB device (Type B) Ionizer

Read more
Show Less

AP224X

Capacity

220 g

mol Conversion Function

Included

Display

OEL display (dot matrix)

Weight

Approx. 7.0 kg

Body Dimensions

Approx. 212(W) x 367(D) x 345(H)mm

Pan Size

ø91 mm

Sensitivity Stability Against Temperature Range

±2 ppm/°C (10 to 30°C)

Operating Temperature/Humidity Range

5 to 40°C 20 to 85% *5

Ionizer

Optional

Clock-CAL

Included

Inspection Support Function

Included

Sample (Concentration) Preparation

Not Included

HPLC Buffer Solution Preparation

Not Included

Minimum display

0.1 mg

Recipe Compounding

Not Included

USB Device (Type B)

Included

USB Host (Type A)

Not Included

Response Time *4 

2 sec.

Response Time for race Measurements *3

2 sec.

Linearity *2

±0.2 mg

Minimum Weight *2

200 mg

Repeatability (Standard deviation)

0.1 mg

External Calibration Weight Range for Span Calibration

95 to 220.009 g
(200 g)

Calibration Weight

Built-in

Input/Output Terminal

RS-232C (D-sub9P plug)
USB device (Type B) Ionizer

Read more
Show Less

AP324X

Capacity

320 g

mol Conversion Function

Included

Display

OEL display (dot matrix)

Weight

Approx. 7.0 kg

Body Dimensions

Approx. 212(W) x 367(D) x 345(H)mm

Pan Size

ø91 mm

Sensitivity Stability Against Temperature Range

±2 ppm/°C (10 to 30°C)

Operating Temperature/Humidity Range

5 to 40°C 20 to 85% *5

Ionizer

Optional

Clock-CAL

Included

Inspection Support Function

Included

Sample (Concentration) Preparation

Not Included

HPLC Buffer Solution Preparation

Not Included

Minimum display

0.1 mg

Recipe Compounding

Not Included

USB Device (Type B)

Included

USB Host (Type A)

Not Included

Response Time *4 

2 sec.

Response Time for race Measurements *3

2 sec.

Linearity *2

±0.3 mg

Minimum Weight *2

200 mg

Repeatability (Standard deviation)

0.15 mg

External Calibration Weight Range for Span Calibration

95 to 320.009 g
(300 g)

Calibration Weight

Built-in

Input/Output Terminal

RS-232C (D-sub9P plug)
USB device (Type B) Ionizer

Read more
Show Less

AP124Y

Capacity

120 g

mol Conversion Function

Not Included

Display

OEL display (dot matrix)

Weight

Approx. 6.5 kg

Body Dimensions

Approx. 212(W) x 367(D) x 345(H)mm

Pan Size

ø91 mm

Sensitivity Stability Against Temperature Range

±2 ppm/°C (10 to 30°C)

Operating Temperature/Humidity Range

5 to 40°C 20 to 85% *5

Ionizer

Not Included

Clock-CAL

Not Included

Inspection Support Function

Not Included

Sample (Concentration) Preparation

Not Included

HPLC Buffer Solution Preparation

Not Included

Minimum display

0.1 mg

Recipe Compounding

Not Included

USB Device (Type B)

Included

USB Host (Type A)

Not Included

Response Time *4 

2 sec.

Response Time for race Measurements *3

2 sec.

Linearity *2

±0.2 mg

Minimum Weight *2

200 mg

Repeatability (Standard deviation)

0.1 mg

External Calibration Weight Range for Span Calibration

45 to 120.009 g
(100 g)

Calibration Weight

No

Input/Output Terminal

RS-232C (D-sub9P plug)
USB device (Type B)

Read more
Show Less

AP224Y

Capacity

220 g

mol Conversion Function

Not Included

Display

OEL display (dot matrix)

Weight

Approx. 6.5 kg

Body Dimensions

Approx. 212(W) x 367(D) x 345(H)mm

Pan Size

ø91 mm

Sensitivity Stability Against Temperature Range

±2 ppm/°C (10 to 30°C)

Operating Temperature/Humidity Range

5 to 40°C 20 to 85% *5

Ionizer

Not Included

Clock-CAL

Not Included

Inspection Support Function

Not Included

Sample (Concentration) Preparation

Not Included

HPLC Buffer Solution Preparation

Not Included

Minimum display

0.1 mg

Recipe Compounding

Not Included

USB Device (Type B)

Included

USB Host (Type A)

Not Included

Response Time *4 

2 sec.

Response Time for race Measurements *3

2 sec.

Linearity *2

±0.2 mg

Minimum Weight *2

200 mg

Repeatability (Standard deviation)

0.1 mg

External Calibration Weight Range for Span Calibration

95 to 220.009 g
(200 g)

Calibration Weight

No

Input/Output Terminal

RS-232C (D-sub9P plug)
USB device (Type B)

Read more
Show Less

AP324Y

Capacity

320 g

mol Conversion Function

Not Included

Display

OEL display (dot matrix)

Weight

Approx. 6.5 kg

Body Dimensions

Approx. 212(W) x 367(D) x 345(H)mm

Pan Size

ø91 mm

Sensitivity Stability Against Temperature Range

±2 ppm/°C (10 to 30°C)

Operating Temperature/Humidity Range

5 to 40°C 20 to 85% *5

Ionizer

Not Included

Clock-CAL

Not Included

Inspection Support Function

Not Included

Sample (Concentration) Preparation

Not Included

HPLC Buffer Solution Preparation

Not Included

Minimum display

0.1 mg

Recipe Compounding

Not Included

USB Device (Type B)

Included

USB Host (Type A)

Not Included

Response Time *4 

2 sec.

Response Time for race Measurements *3

2 sec.

Linearity *2

±0.3 mg

Minimum Weight *2

200 mg

Repeatability (Standard deviation)

0.15 mg

External Calibration Weight Range for Span Calibration

95 to 320.009 g
(300 g)

Calibration Weight

No

Input/Output Terminal

RS-232C (D-sub9P plug)
USB device (Type B)

Read more
Show Less

AP135W

Capacity

135 g

Minimum display

0.01 mg

{"title":"Specification Sheet Downloads","description":"Download the latest specification sheets.","source":"product","key":3525,"max":30,"filter_types":["spec_sheet"],"link_title":"View other Downloads","link_url":""} {"title":"Downloads","description":"Download the latest brochure.","source":"product","key":3525,"max":30,"filter_types":["brochures"],"link_title":"View other Downloads","link_url":"","pdf_links":[]}