Error Code | Description |
---|---|
ERR NO CONST. PRS | Description: A constant pressure pumping mode is set on the system controller with [1: 20A] set at [OP MODE]. Resolution: Use the constant flow pumping mode. |
ERR NO LPGE MODE | Description: A constant pressure pumping mode is set on the system controller with [1: 20A] set at [OP MODE]. Resolution: Select another control mode. |
ERR FLOW TOO HI |
Description: A flow rate higher than 5mL/min is set on the system controller with [1: 20A] set at [OP MODE]. |
WARN : LOST PARAM |
Description: Parameters may have been lost. |
SUB ROM FAILURE | Description: ROM error (electric failure) Resolution: Turn power OFF and contact your Shimadzu representative. |
MAIN ROM FAILURE | Description: ROM error (electric failure) Resolution: Turn power OFF and contact your Shimadzu representative. |
0x2F80 | Description: Undefined warning. |
0x00D0 | Description: Invalid value. |
0x8568 | Description: Rinse liquid is low. (Autosampler) |
0x80F7 | Description: PDA Detector is too busy. (PDA Detector) |
0x80F5 | Description: Connection cancel because of changing PDA detector's group name/system name. (PDA Detector) |
0x80F3 | Description: Connection cancel because of changing PDA detector's IP address. (PDA Detector) |
0x80F0 | Description: Connection request received from more than one PDA detector. (PDA Detector) |
0x80E0 | Description: The system which can be registered on the group is 10 in the maximum. |
0x8084 | Description: PDA detector cannot accept communication command. (PDA Detector) |
0x8002 | Description: Confirm startup time. |
0x8001 | Description: System check timeout error has occurred. |
0x4004 | Description: Buffering error. |
0x4001 | Description: Buffering error. |
0x2F3F | Description: Communication error. |
0x2F10 | Description: Communication error. |
0x2F0A | Description: OptLink memory error. |
0x2F01 | Description: [0x2F01 to 0x2F0A] OptLink memory error. |
0x89FB | Description: Undefined warning. |
0x89D2 | Description: Temperature is increasing inside the instrument. |
0x8954 | Description: LAN timeout error has occurred. (CBM) |
0x8952 | Description: LAN timeout error has occurred. (PC) |
0x8721 | Description: Low set temperature. |
0x87D0 | Description: Low set temperature. |
0x0050 | Description: No scan data |
0x0049 | Description: Auto zero level is out of range. Confirm the installation of the cell, no bubble in the cell, and the settings of SENS and GAIN parameters. |
0x0046 | Description: Operation mode is local. Set the operation mode to remote. |
0x8632 | Description: Cannot write in CMD. |
0x0078 | Description: High set temperature. |
0x0077 | Description: Low set temperature. |
0x0075 | Description: Operation mode is local. Set the operation mode to remote. |
0x85F1 | Description: Full rack is detected. Remove the duplex rack on the stack or set the dummy rack. If the rack is not changed, the sequence will be stopped. |
0x85F0 | Description: The reserved warning code 1 has been detected. |
0x85EF | Description: No rack is detected. Check the rack on the stack. If the rack is not set and analysis is executed, the sequence will be stopped. |
0x85EE | Description: As the stack has been closed, the temperature control has been on again. |
0x85ED | Description: The temperature control has been off because stack has been opened for more than 30 minutes. If the stack is closed, the temperature control will be on again. |
0x85EC | Description: The stack is open. Close the stack in the rack-changer soon. |
0x85EB | Description: It is necessary to set all racks on the stacks at the rackchanger. Check all racks are set. |
0x8548 | Description: The temperature control of rack plate R has been turned off because no rack has been set for more than 30 minutes. If the rack is set, the temperature control will be on again. |
0x8547 | Description: The temperature control of rack plate M has been turned off because no rack has been set for more than 30 minutes. If the rack is set, the temperature control will be on again. |
0x8546 | Description: The temperature control of rack plate L has been turned off because no rack has been set for more than 30 minutes. If the rack is set, the temperature control will be on again. |
0x8545 | Description: Racks with same rack ID are set. |
0x8544 | Description: Invalid rack ID. |
0x8543 | Description: Rack on rack plate R is not adjusted. |
0x8542 | Description: Rack on rack plate M is not adjusted. |
0x8541 | Description: Rack on rack plate L is not adjusted. |
0x0094 | Description: Operation mode is local. Set the operation mode to remote. |
0x81F1 | Description: A seal malfunction has occurred in the online degassing unit vacuum line. |
0x81E7 | Description: Cannot write parameters in EEPROM. |
0x81D3 | Description: Mobile phase D is low. |
0x81D2 | Description: Mobile phase C is low. |
0x81D1 | Description: Mobile phase B is low. |
0x81D0 | Description: Mobile phase A is low. |
0x0033 | Description: Operation mode is local. Set the operation mode to remote. |
0x0006 | Description: Illegal backup parameter. System information and user information have been changed. Resolution: Some information has been initialized. Check the system and user information. |
0x0005 | Description: Illegal backup parameter. Method and sequence files have been changed. Resolution: Some parameters have been initialized. Check the methods and sequences. |
0x0004 | Description: Data not protected. Method and sequence files have been restored. Resolution: The method and sequence files have been returned to the state they were in the previous time the system controller was started up. Check the methods and sequences. |
0x0003 | Description: Data not protected. Method and sequence files have been initialized. Resolution: Method and sequence files have been initialized. After making the appropriate settings for the connected components in the "Configuration" tab page, reset the methods and sequences in all the relevant windows. |
ROM FAILURE | Description: ROM error (electric failure) Resolution: Turn power OFF and contact your Shimadzu representative. |
ERROR HOME POS | Description: The motoro home position cannot be detected, the motor does not run, or the motor is slipping. Resolution: Turn the power OFF and ON again, and press PUMP. If this message still appears, trun the power OFF and contact your Shimadzu representative. |
WARN : M.PHASE | Description: The remaining volume of mobile phase is less than the alarm level. Resolution: Replace the mobile phase and reset the amount of mobile phase./td> |
WARN :LOST PARAM | Description: Parameters may have been lost. Resolution: Parameter resetting is required. Contact your Shimadzu representative. |
WARN : DEGAS PRS | Description: Vacuum pressure of degassing unit (DGU-20A) connected to this instrument has been abnormal for a certain time. Resolution: Check the vacuum pressure of degassing unit. If the vacuum pressure is [NG], turn the power OFF once and turn it ON again. If the vacuum pressure is [OK], it is of no matter, for the degassing unit is returning to run automatically |
WARN : M.PHASE-B | Description: The remaining volume of pump A's (or pump B's) mobile phase is less than the alarm level. Resolution: Replace pump A's (or pump B's) mobile phase and reset the amount. |
WARN : M.PHASE-A | Description: The remaining volume of pump A's (or pump B's) mobile phase is less than the alarm level. Resolution: Replace pump A's (or pump B's) mobile phase and reset the amount. |
ERROR EXTERNAL | Description: An error signal was sent from the external device connected to the external input/output terminal [ERR IN]. Resolution: Inspect the external device, and eliminate the cause of the error |
ERROR LEAK | Description: Leak sensor has detected leakage. Resolution: Inspect the leakage in the flow line. Wipe away any liquid around the leak sensor. |
OPEN DRAIN VALVE | Description: Pump discharge pressure has exceeded [P-PMAX] during purging. Resolution: Open the drain valve before pressing PURGE . If an error message appears during auto-purging by system controller, inspect the clogging in the flow line. If no clogging is found, set the appropriate [P-FLOW] and [P-PMAX] value. |
ERROR B-HOME POS | Description: Pump A's (or pump B's) motor home position cannot be detected, the motor does not run, or the motor is slipping. Resolution: Turn the power OFF and ON again, and press PUMP . If this message still appears, turn the power OFF and contact your Shimadzu representative. |
ERROR A-HOME POS | Description: Pump A's (or pump B's) motor home position cannot be detected, the motor does not run, or the motor is slipping. Resolution: Turn the power OFF and ON again, and press PUMP. If this message still appears, turn the power OFF and contact your Shimadzu representative. |
ERROR P-MIN | Description: Pump discharge pressure has fallen below the set min. pressure limit, [P.MIN]. Pumping will automatically stop Resolution: Inspect the leakage in the flow line. If no leakage is found, set the appropriate P.MIN value. Note that [P.MIN] function is not activated for first one minute of pumping. |
ERROR P-MAX | Description: Pump discharge pressure has exceeded the set max. pressure limit, [P.MAX]. Pumping will automatically stop, unless [S-PROT] function has been activated, in which case pumping will continue at half flow rate Resolution: Inspect the clogging in the flow line. If no clogging is found, set the appropriate [P.MAX] value. |
OUT OF MAXSTEPS | Description: More than 320 steps of time program have been set. Resolution: Delete unnecessary programs. |
A-B TRANS ERROR | Description: A transmission error has occurred between pumps A and B. Resolution: Turn power OFF and contact your Shimadzu representative. |
ERROR OVER HEAT | Description: Interior temperature has risen to an abnormal level. Resolution: Check to be sure that the internal fan can move, and that the rear exhaust vent is not blocked. If this message still appears, turn the power OFF and contact your Shimadzu representative. |
RAM FAILURE | Description: RAM error (electric failure) Resolution: Turn power OFF and contact your Shimadzu representative. |