Troubleshooting Reproducibility Problems
Nexera Virtual Advisor
Symptom | Cause | Solution |
---|---|---|
|
Recorder pen is below baseline | Press [Zero] to return the pen to the baseline. |
|
Varying mobile phase composition | Examine the pump functionality and take corrective action as necessary. Use a fresh mobile phase preparation and check the composition of the new mobile phase. |
Poor reproducibility | Needle seal has deteriorated | Replace the needle seal. |
|
Leak - flow lines | Find and repair the leak. |
|
Insufficient flow line rinse or no rinse solution | Increase the RINSE VOL settings or replenish the rinse solution. Perform the PUMP HEAD FLUSH operation. PURGE or RINSE the flow lines. |
|
Fluctuating ambient temperature | Install unit in a room with minimal temperature variation. |
Poor reproducibility | Column performance has deteriorated | Replace column. |