Troubleshooting Peak Problems
Nexera Virtual Advisor
Symptom | Cause | Solution |
---|---|---|
|
Varying mobile phase composition | Use a fresh mobile phase preparation and check the composition of the new mobile phase. |
Retention time fluctuation | Unstable flow rate | Examine the pump function and take corrective action as necessary. |
Retention time fluctuation | Fluctuating ambient temperature | Relocate the instrument or protect the current location from excessive environmental effects. Use a column oven. |
Retention time fluctuation | Column performance has deteriorated | Examine the column oven function and take corrective action as necessary. Replace the column. |
Retention time fluctuation | Check valve is not working properly | Clean the check valve in an ultrasonic bath or replace as necessary. Pump 2-propanol through the check valve to clean it. |
Retention time fluctuation | Blockage present in the needle or plumbing | Clean the high-pressure valve with reverse cleaning of the flow lines. Remove the blockage with reverse cleaning of the flow lines. |
|
No plate in the specified rack position | Insert a plate with samples in the specified rack position. |
|
Needle contacts the bottom of the plate or misses the target | Determine the stack application (microtiter plate or deep-well microfilter plate) and set the needle stroke accordingly. Perform rack teaching for the SIL-20A/20AC, if applicable. |
|
Flow line blocked with plate material particles | Remove the blockage with reverse cleaning of the flow lines. |
|
No mobile phase flow | Examine the pump function and take corrective action as necessary. |
|
Insufficient sample in vials | Add more sample to the vials. |
|
Detector is not connected or is faulty | Examine the detector connections and take corrective action as necessary. |
No peaks | Column performance has deteriorated | Replace the column. |
No peaks | Blockage in the injection flow lines | Check and correct the content of the user-written program. Clear or replace any blocked plumbing. |
|
Leak - flow lines | Find and repair the leak. |
|
Dead volume between the flow line connections | Examine flow line connections. Reconfigure the connections to eliminate any dead volume. |
|
Connections between the pump and column are reversed | Reconfigure the plumbing connections. |
|
Column performance has deteriorated | Replace the column. |
|
Recorder pen is below the baseline | Press [Zero] to return the pen to the the baseline. Remove the blockage with reverse cleaning of the flow lines. Replace the needle or the plumbing. |
|
Varying mobile phase composition | Examine the pump functionality and take corrective action as necessary. Use a fresh mobile phase preparation and check the composition of the new mobile phase. |
|
Needle seal has deteriorated | Replace the needle seal. |
|
Leak - flow lines | Find and repair the leak. |
|
Insufficient flow line rinse or no rinse solution | Increase the RINSE VOL settings or replenish the rinse solution. Perform the PUMP HEAD FLUSH operation. PURGE or RINSE the flow lines. |
|
Fluctuating ambient temperature | Install unit in a room with minimal temperature variation. |
|
Column performance has deteriorated | Replace column. |