- READY indicator will not illuminate
|
Oven indicator not illuminated |
Press the Oven key to illuminate the oven indicator. Execute a 'run' signal from the system controller. |
- READY indicator will not illuminate
|
COOLER MODE set to 0 |
Set the COOLER MODE parameter to 1 or 2. |
|
Power supply specifications have not been met |
Use a power supply that meets the instrument specifications. |
|
Power cable is unplugged |
Properly connect the power cable to the power supply. |
|
Power cable is cut or broken |
Replace the power cable with a new one. |
- Leak - oven (no error message)
|
Leak sensor sensitivity is incorrectly adjusted |
Readjust the leak sensor sensitivity. |
|
Display key has not been pressed |
Press the Display key to illuminate the operation keys and restore keypad function. |
- COOLER MODE is ON, but the temperature does not fall to set the temperature
|
Oven interior is not properly sealed |
Ensure that oven door is closed and tubing insulating grommets are in place. |
- COOLER MODE is ON, but the temperature does not fall to set the temperature
|
Insufficient clearance around the oven |
Allow at least 200 mm behind and at least 20 mm on both sides of the oven. |
- COOLER MODE is ON, but the temperature does not fall to set the temperature
|
High ambient temperature |
Eliminate the cause of the high temperature. |