Troubleshooting Baseline Problems
Nexera Virtual Advisor
Symptom | Cause | Solution |
---|---|---|
Drift | Dirty cell lenses | Dismantle the flow cell and clean or replace the lenses as necessary. |
Drift | Dirty flow lines | Thoroughly clean flow lines. |
Drift | Faulty detector | Check the detector function and take any necessary corrective action. |
Drift | Fluctuating ambient temperature | Install the unit in a room with minimal temperature variation. |
Drift | Fluctuating flow rate | Check the pump function and take any necessary corrective action. |
Drift | Impurities in the mobile phase - drift stops when the pump is off or an air cell is used | Inspect the mobile phase and clean the flow lines. |
Drift | Lamp intensity has deteriorated - noise continues when an air cell is used | Replace the lamp. |
Drift | Lamp not stabilized | Start measurement 1 to 1.5 hours after the lamp was turned on. |
Drift | Lamp or flow cell is improperly installed | Reinstall the lamp or flow cell. |
Drift | Leak - detector | Examine the cell tubing and cell window for leaks. Clean and repair the leak. |
Drift | Mobile phase contains impurities or the column has not equilibrated | Examine or prepare fresh mobile phase. Allow more time for column equilibration. |
Drift | Strong air current or large ambient temperature changes | Relocate the instrument, or protect the current location from excessive environmental effects. |
Drift and Noise | Bubbles in flow cell | Perform the bubble removal procedure. |
Fluctuation | Strong air current or large ambient temperature changes. | Relocate the instrument, or protect the current location from excessive environmental effects. |
No Change | Faulty circuits | Replace any faulty parts. |
No Change | Lamp is off | Set the LAMP parameter to 1 or 2 to turn ON lamp power. |
No Change | RANGE set to 0 | Enter an appropriate RANGE value. |
Noise | Cell lenses are dirty | Dismantle the flow cell and clean or replace the lenses as necessary. |
Noise | External vibrations | Reduce the external vibrations. Use a sturdy counter as a support. |
Noise | Impurities in the mobile phase - drift stops when the pump is off or an air cell is used | Inspect the mobile phase and clean the flow lines. |
Noise | Lamp intensity has deteriorated - noise continues when an air cell is used | Replace the lamp. |
Noise | Low lamp energy | Replace the lamp. |
Noise | Strong air current or large ambient temperature changes | Relocate the instrument, or protect the current location from excessive environmental effects. |
Noise - two-wavelength mode | Absorbances of mobile phase and sample solvent are different | Select a wavelength with less absorbance or select solvents with closer absorbances. |
Noise - two-wavelength mode | Response value is too low | Set the response to a higher value. |
Noise - with pump stroke | Pump pulsation | Install a pump pulsation damper. |
OVER is displayed | Recorder pen is below baseline | Press [Zero] to return the pen to the baseline. |
Sawtooth | Bubbles in flow cell | Apply back pressure to cell outlet with a backpressure device or 0.3 mm I.D. x2 m tubing. Degas the mobile phase. Perform bubble removal procedure. |
Spikes - continuous | Bubbles in flow cell | Apply back pressure to cell outlet with a backpressure device or 0.3 mm I.D. x 2 m tubing. De-gas the mobile phase. |
Spikes - random | Bubbles in the flow cell | Apply back pressure to cell outlet with a backpressure device or 0.3 mm I.D. x 2 m tubing. De-gas the mobile phase. Perform bubble removal procedure. |
Spikes - regular | Bubbles in flow cell | Apply back pressure to cell outlet with a backpressure device or 0.3 mm I.D. x 2 m tubing. Perform the bubble removal procedure. Rinse the flow cell with 2-propanol. Use the supplied syringe. |
Spikes - regular | Pump pulsation | Examine the pump function and take corrective action as necessary. Install a pump pulsation damper. Perform the bubble removal procedure. |
Spikes - with pump stroke | Bubbles in the flow cell | Apply back pressure to cell outlet with a backpressure device or 0.3 mm I.D. x 2 m tubing. Rinse the flow cell with 2-propanol. Use the supplied syringe. |
Swell | Dirty cell lenses | Dismantle the flow cell and clean or replace the lenses as necessary. |
Swell | Impurities in the mobile phase - drift stops when pump is off or the air cell is used | Inspect the mobile phase and clean the flow lines. |
Swell | Lamp intensity has deteriorated - noise continues when air cell is used | Replace the lamp. |
Unstable | Bubbles in the flow cell | Apply back pressure to cell outlet with a backpressure device or 0.3 mm I.D. x 2 m tubing. Rinse the flow cell with 2-propanol. Use the supplied syringe. |