Troubleshooting Autosampler Problems
Nexera Virtual Advisor
Symptom | Cause | Solution |
---|---|---|
|
Rotor and stator seal and worn | Replace the rotor and stator as necessary. |
|
Rotor and stator seal and worn | Replace the rotor, inspect and replace the stator as necessary. |
|
Connections are loose or faulty | Tighten or replace the male nuts and ferrules of the flow line connections. |
|
The flow rate of the mobile phase is too high | Replace the SIL outlet tubing with 0.25 mm ID PEEK tubing. |
|
Incorrect high-pressure valve position - NO HPV HOME displayed | Contact your Shimadzu representative. |
|
High-pressure valve blockage | Disassemble and clean the high-pressure valve. Replace the rotor and stator as necessary. |
|
Flow line blockage | Clean or replace any blocked flow lines. Remove the blockage with reverse cleaning of the flow lines. |
|
Flow line blockage | Clean or replace any blocked flow lines. Remove the blockage with reverse cleaning of the flow lines. |
|
Column blockage | Replace the column. |
|
Power supply specifications have not been set | Use a power supply that meets the instrument specifications. |
|
Power cable is unplugged | Properly connect the power cable to the power supply. |
|
Power cable is cut or broken | Replace the power cable with a new one. |
|
Fuse is blown | Replace the fuses. If fuses blow out frequently, contact your Shimadzu representative. |