hplc-knowledge-base-optimize

Just for Fun – Explain the Difference

Look at the three chromatograms below.

  • They are all performed with “C18” column chemistries.
  • The same 8 component test mix was injected on each column.
  • All 3 separations are performed using the same mobile phase.
  • All 3 columns have the same hardware format (length, width).
  • All 3 columns are packed with the same particle size stationary phase.

How can it be? How can the results be so different?

nexera-hplc-optimize-c18-3

 

1. Uracil

2. 3-butylpyridine

3. Phenol

4. 4-phenylbutyric acid

5. N,N-diethyl-meta-toluamide

    (DEET)

6. Quinizarin

7. Propylbenzene

8. Butylbenzene

Mobile Phase:  25mM KH2PO4

  pH 2.5:ACN

  (35:65)

Flow Rate:  1.0mL/min

Detector:  UV at 254nm

The answer lies in the bonded phase chemistry of the packed bed.  The columns each have a different % carbon loading (bonded phase coverage on the silica) and a different kind of end-capping.  Vastly different results. Proof that a C18 is not a C18 is not a C18.

just-for-fun-c18
just-for-fun-c18-b
just-for-fun-c18-c

 

General Recommendations

  • Plumb systems with 0.005” tubing or smaller ID
  • Keep injected volumes between 1 and 5µL (~1% of void volume)
  • Always use high quality HPLC-grade or UHPLC-grade mobile phase solvents.
  • Consider pre-filtering mobile phase solvents through 0.45µm filters, especially after modification with a buffer
  • Always pre-filter samples through 0.45µm or 0.2µm syringe filters or use filter vials that provide the same function.
  • Consider a 100 x 3mm pack with 2.7 μm SPP particles for quality HPLC
    • higher efficiency than standard HPLC
    • lower back pressure than 3 μm traditional particles
    • more resistant to clogging than traditional particles
  • Consider a 50 x 2.1mm packed with sub-2µm particles optimal for UHPLC
    • comparable theoretical plates, N, to a 150 x 4.6mm, 5µm
    • optimal flows in the range 300 – 600µL/min
    • run times from 3 – 5min. (100 x 2.1mm, 7 – 10min.)
    • back pressure will be higher

Contributions to band broadening

contributions-to-band-broadening
Peak broadening occurs due to contributions from the intra-column volume and extra-column volume –
  • Injector mechanism
  • Detector flow cell
  • Connection tubing between injector and column
  • Connection tubing between column and detector flow cell

The smaller the intra-column volume, the higher the contribution from extra-column volume

Good UHPLC performance demands that we minimize the system volume by using minimum tubing IDs and the shortest lengths.

Animation Explanation

band-broadening

 Too much extra-column volume will result in peak broadening. This is called dispersion. Perceived chromatographic resolution will be lower and detector sensitivity will be lower because of reduced peak height.

Optimizing for Delay Volume

optimizing-delay-volume


Vd = Vmixer + Vinj + Vtubing

Gradient Delay volume is the total amount of volume from the point of mixing to the head of the column.

Gradient Delay or “Dwell” Volume

Gradient delay is most noticeable with small-volume LC columns. As a practical example consider a 100 x 2.1mm column operating at a flow rate of 200 μL/min on a system with 200 μL of delay volume. It would take 1 minute for the column to “see” the gradient –the first minute would be initial isocratic condition. That “1 minute” may be 20% or more of the run time.

Compensating for Gradient Delay

 

compensating-gradient-delay

Practical Method Considerations

Title Void Volume Eq. Time @400µL/min.
50 x 2.1mm, 1.9µm 0.1mL 0.25min
100 x 2.1mm, 1.9µm 0.2mL 0.5min

Sample Matrix Strength

 

MP A: Water

MP B: Acetonitrile

Gradient: 10 – 100% B over 5 minutes

Sample Matrix: 100% Acetonitrile

MP A: Water

MP B: Acetonitrile

Gradient: 5 – 100% B over 5 minutes

Sample Matrix: 10% Acetonitrile in Water

sample-matrix-strength-1 sample-matrix-strength-2

Some Basic HPLC Flow Principles

optimizing-chrom-basic-hplc
optimizing-chrom-basic-hplc-diffusion
optimizing-chrom-basic-hplc-long-diffusion

Detector Response and Column ID

Reducing the column ID ( ↓ radial diffusion ) increases the peak height/sensitivity:

 

optimizing-chrom-detector-response

Solvent Consumption and Column ID

Column format Solvent Consumed in five 20min. Runs* Reduction in solvent used
4.6 x 150mm 100mL 0
3.0 x 150mm 42mL 58%
2.1 x 150mm 13mL 87%


* Flow rate = 1mL/min. 

Choosing HPLC/UHPLC Columns

Choose columns according to analysis goals (sample throughput, resolution, target chemistries)

 

 

SPP – Superficially Porous Particles

 

spp-image

2.7µm Superficially Porous Particles offer high resolution and fast separations approaching that of UHPLC but without the higher backpressures.

spp-chromatogram1

spp-chromatogram2

 

 

Isocratic vs. Gradient Elution Modes

Isocratic Elution

Gradient Elution

 

Method Development – Suggested Screening Gradient

Recommended LC Column:

method-development
Shimadzu LabSolutions Pump Program – Screening Gradient

 

Screening Gradient

screening-gradient

 

Outcome Scenario #1: Gradient Elution (Weak “A” – Strong “B” solvent system)

screening-gradient-outcome-scenario

Interpretation

For a reversed phase LC column where A is water and B is organic, this first run shows sample constituents that are all polar. So, the separation is poor because the substances experience little partitioning on the stationary phase. In other words, the weak, starting solvent condition brings the sample constituents off too early.

 

Outcome Scenario #1: Gradient Elution (Weak “A” – Strong “B” solvent system)

screening-gradient-outcome-scenario

Interpretation

To cause the substances to partition more into the stationary phase, we reduce the slope of the gradient so that the mobile phase strength does not increase as quickly. The substances retain longer and begin to separate from one another. Notice that we still “flush” the column with strong solvent at the end of the run to effectively clean the column.

 

Outcome Scenario #1: Gradient Elution (Weak “A” – Strong “B” solvent system)

screening-gradient-outcome-scenario

Interpretation

We reduce the gradient slope again until we begin seeing a good separation within the time frame.

 

Outcome Scenario #2: Gradient Elution (Weak “A” – Strong “B” solvent system)

screening-gradient-outcome-scenario

Interpretation

For a reversed phase LC column where A is water and B is organic, this first run shows sample constituents that are all very non-polar. So, the separation is poor because the substances come off late –only when the solvent strength is high.

 

Outcome Scenario #2: Gradient Elution (Weak “A” – Strong “B” solvent system)

screening-gradient-outcome-scenario2-gradient

Interpretation

Here, we start the gradient run at a higher solvent strength so the sample constituents begin partitioning into the stationary phase earlier.

 

Outcome Scenario #2: Gradient Elution (Weak “A” – Strong “B” solvent system)

screening-gradient-outcome-scenario2-gradient

Interpretation

Again, the gradient is adjusted to start at a higher solvent strength so the sample constituents begin partitioning into the stationary phase earlier.

 

 

nexera-e-book-banner

 

Back to Index