Q.What kind of balance is needed to build a system compliant with guidelines and standards such as GLP, GMP and the ISO 9000 Series?
A : To conform to these standards, the balance to be used must be already adjusted to within standards predetermined in your own departments, periodically adjusted and a record of the results of inspection must be kept.
Inspection items that must be performed on a daily basis include external appearance (e.g. dirt) on the balance body and display, horizontality and sensitivity checks (calibration).
- Sensitivity check can be easily conducted if you have an electronic balance with built-in calibration weights.
There are also models that incorporate functions such as Perfect Self Calibration (PSC), Clock-CAL that automatically perform calibration for you. - For inspections, a record clearly indicating the date/time that inspection was conducted must be kept. Models with built-in ISO print function can be connected to a printer to provide a date/time-stamped balance calibration record.
Use of balances with these functions is very convenient.
Shimadzu also conducts JCSS calibration services for customers' own weights and balances.